Title: Karakterisasi 2D Ion Chamber Array Seven29 pada Pesawat Linear Accelerator (LINAC) Elekta CompactTM
Authors: Nainggolan, Bill Samuel
Keywords: radioterapi;Treatment Planning System;QA Plan
Issue Date: 2019
Publisher: Program Studi Fisika FSM-UKSW
Abstract: Dalam penelitian ini kami mengevaluasi karakteristik ruang 2D array dan kinerjanya dalam verifikasi rencana radioterapi. 2D Array yang digunakan adalah model 2D Array Octavius Detector Seven29 (T10040), detektor array 2D mempunyai 729 ruang ionisasi diatur secara seragam dalam matriks 27×27 cm2 dengan area aktif 27×27 cm2. Reproduksibilitas jangka menengah detektor telah diuji melalui serangkaian pengukuran berulang. Reproduksibilitas jangka menengah berada dalam 0,67%. Linearitas dosis juga diuji, respon sistem terhadap dosis diverifikasi menjadi 3, 5, 10, 20, 50, 100, 150, 200, 250 dan 300 MU. Faktor keluaran sesuai dan sangat baik pada pengukuran bilik, dengan selisih 0,03971, sensitivitas 2D Array standar devisiasi maksimum 0,908%. Ketika digunakan untuk mengevaluasi planning dengan hasil diatas 97% batas klinis. Pengujian yang dilakukan dalam penelitian ini menunjukkan bahwa 2D Array Seven29 adalah dosimeter yang sesuai dan akurat dan dapat menjadi alat yang berguna untuk jaminan kualitas dan verifikasi rencana radioterapi.
In this study we evaluated the characteristics of 2D array space and its performance in verifying radiotherapy plans. 2D Arrays used are 2D Array Octavius Detector Seven29 models (T10040), 2D array detectors have 729 ionization spaces arranged uniformly in a 27×27 cm2 matrix with an active area of 27×27 cm2. The medium-term reproducibility of the detector has been tested through a series of repeated measurements. Medium-term reproducibility is 0.67%. The dose linearity was also tested, the system response to the dose was verified to be 3, 5, 10, 20, 50, 100, 150, 200, 250 and 300 MU. The output factor is suitable and very good in the measurement chamber, with a difference of 0.03971, the sensitivity of the 2D Array standard deviation is a maximum of 0.908%. When used to evaluate planning with results above 97% clinical limits.. The tests conducted in this study indicate that 2D Array Seven29 is a suitable and accurate dosimeter and can be a useful tool for quality assurance and verification of radiotherapy plans.
Description: Tidak diijinkan karya tersebut diunggah ke dalam aplikasi Repositori Perpustakaan Universitas karena telah proses publikasi Jurnal Nasional Terakreditasi .
URI: https://repository.uksw.edu/handle/123456789/19559
Appears in Collections:T1 - Physics

Files in This Item:
File Description SizeFormat 
T1_642014021_Abstract.pdfAbstract320.06 kBAdobe PDFView/Open
T1_642014021_Full text.pdf
  Restricted Access
Full text1.66 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.