Title: Validasi Metode dan Penetapan Kadar Silika dalam Air Drum Pembangkit Listrik Tenaga Gas & Uap Priok
Other Titles: Validation Methods and Determination of Silica Content in Water Drum Gas and Steam Power Plant Priok
Authors: Oliver, Lorenzo
Keywords: ANSA;kadar silika;spektrofotometri UV-Vis;validasi metode
Issue Date: 2019
Publisher: Program Studi Kimia FSM-UKSW
Abstract: Silika dalam air umpan Pembangkit Listrik Tenaga Gas dan Uap (PLTGU) mampu menimbulkan kerak yang menyebabkan gangguan pada perputaran turbin serta kerusakan alat. Penetapan kadar silika yang dilakukan secara rutin memerlukan reagen yang bergantung akan ketersediaanya. 4-amino-2-naphthol-1-sulphonic acid (ANSA) digunakan sebagai reagen alternatif untuk menunjang ketersediaan reagen set. Validasi metode perlu dilakukan untuk menjamin bahwa metode analisis dengan reagen alternatif telah sesuai dan dapat diandalkan. Penelitian ini bertujuan untuk menetukan validasi metode analisis kadar silika dengan metode spektrofotometri serta menerapkan metode yang dikembangkan untuk penetapan kadar silika dalam air drum PLTGU. Metode yang digunakan untuk penetapan kadar silika adalah spektrofotometri UV-Vis. Berdasarkan hasil yang diperoleh, panjang gelombang penetapan kadar silika adalah 816 nm meliputi persamaan pada kurva baku silika y=0,0014x-0,0594 dengan koefisien determinasi sebesar 0,0995. Pengujian akurasi menggunakan tes recovery adalah 77,14-95,29% pada konsetrasi 90, 115, dan 155 ppb. Pengujian presisi intraday dan interday masing-masing pada kisaran 0,38-10,61% dan 0,45-14,80%pada konstrasi yang sama dengan pengujian akurasi. Nilai LOD dan LOQ adalah 2,48 ppb dan 8,28 ppb. Pengujian robustness dapat diterima sampai dengan retensi waktu 60 menit . Pengujian stabilitas reagen dapat diterima dengan penambahan reagen ANSA sampai dengan 4 ml. Konsentrasi kadar silika dari air drum LP 3.1, IP 3.1, dan HP 3.1 masing- masing sebesar 302,62;460,55; dan 91,95 ppb.
Silica in the feed water of the Gas and Steam Power Plant (PLTGU) is able to cause scale which disruption to the turbine rotation and damage to the equipment. Determination of silica levels carried out routinely requires the reagents needed to be available. 4-amino-2-naphthol-1-sulphonic acid (ANSA) is used as an alternative reagent to support an alternative set of reagents. A validation method needs to be carried out to ensure that the analysis method with alternative reagents is appropriate and reliable. This study aims to determine the validation of the silica content analysis method with the spectrophotometric method and apply the method developed for the determination of silica levels in the PLTGU air drums. The method used for determination of silica levels is UV-Vis spectrophotometry. Based on the results obtained, the wavelength of determining silica content is 816 sharing the equation on the standard curve y = 0.0014x-0.0594 with a coefficient of determination of 0.0995. Accuracy testing using recovery tests was 77.14-95.29% at 90, 115, and 155 ppb concentrations. Intraday and interday precision testing respectively in the range of 0.38-10.61% and 0.45-14.80% in the same construction with accuracy testing. LOD and LOQ values are 2.48 ppb and 8.28 ppb. Robustness testing can be received up to 60 minutes retention time. The test can be approved with ANSA-approved reagents up to 4 ml. Silica concentration of LP 3.1, IP 3.1 and HP 3.1 air drums respectively 302.62; 460,55; and 91.95 ppb.
Description: Tidak diizinkan karya tersebut diunggah ke dalam aplikasi Repositori Perpustakaan Universitas.
URI: https://repository.uksw.edu/handle/123456789/20684
Appears in Collections:T1 - Chemistry

Files in This Item:
File Description SizeFormat 
T1_652015020_Abstract.pdfAbstract89.79 kBAdobe PDFView/Open
T1_652015020_Full text.pdf
  Restricted Access
Full text2.07 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.